Determination of Assay for Dibenzo-18-Crown-6
Aug 16, 20221. Reagents and Solutions
Acetonitrile: HPLC grade
Water: Ultra pure water
2. Instruments
Liquid chromatograph: UV detector with variable wavelength
chromatographic column: 150mm×4.6mm×5µm Kromasil C18
Microinjector: 50 μL
3. Operating conditions of liquid chromatography
Mobile phase: Acetonitrile+Water=60+40 (Volume ratio);
Rate of flow: 1.0ml/min;
Column temperature: 40℃;
Detection wavelength: 254nm;
Injection volume: 20μL
Retention time: Dibenzo 18-Crown-6 about 4min;
Quantitative method: Area percentage method.
4. Determination steps
4.1 Calculation
Assay of Dibenzo 18-Crown-6 Xi (%), calculate as formula:
In formula:
Xi: Percentage content of Dibenzo 18-Crown-6;
Ai: Peak area of Dibenzo 18-Crown-6;
∑Ai: Sum of chromatographic peak areas in sample solution.
4.2 Allowable difference
The difference between the two parallel determinations results should not exceed 0.50%, taking the arithmetic mean as the determination result.